HI-3010 Filler/Dispenser/IBC

102

Service Manual

C1 - Guidelines for Instabilities on formerly operating systems with Smart Diagnostics

TEST

COMPLETE

Stable?

Yes

No

C1

Repeat for each sensor.

Record load sensor for
comparison and stability

Check individual load sensors output by
using the Smart Diagnostics and
SD summing card

A load sensor output can be considered stable if
the readings only vary +- 0.01mV. Applying
weight and releasing the weight should show a
very rapid Millivolt change. A slow Millivolt
change can indicate a damaged strain guage.

Inspect the summing junction box for
contamination or damage, and replace if necessary

Stable?

Stable?

Insure the problem is not
mechanical. (Review Section B)

If you are unable to isolate the
problem, contact Hardy Technical
Support: 800-821-5831

Yes

No

No

No

Insure the C2 wires are correctly terminated
even if you do not have C2 load sensors.
The IT communications and testing is
conducted over the C2 lines.

Yes

Remove and replace any load
sensor determined to be unstable