Installing the smart diagnostics (-sd) card, Main controller board installed with rear, Rear plate fasteners – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 32: Standoff locations

Advertising
background image

HI-3010 Filler/Dispenser/IBC

16

Service Manual

Place the Main Board rear plate so that the
threaded holes on each side of the instrument
chassis are aligned.

Screw a panhead screw (#4-40) into the
threaded hole on the instrument chassis. Do
not tighten.

Screw the panhead screws that attach the
rear plate to the Main Board until they are
finger tight.

Use a Phillips head screw driver and tighten
all the installed screws until snug.

CAUTION: D

O

NOT

OVERTIGHTEN

.

FIG. 3-13 MAIN CONTROLLER BOARD

INSTALLED WITH REAR PLATE

Step 5. Installation of all the PC Boards used in any HI

3000 Series Instrument requires the same proce-
dures.

Installing the Smart Diagnostics (-SD)
Card

NOTE:

For Configuration Instructions go to Chapter 4 -
Options Configuration/Smart Diagnostics Con-
figuration.

Step 1. Unplug the power connector and cable.
Step 2. Remove the Secure Memory Module.
Step 3. Unplug the load cell connector and cable.
Step 4. Disconnect all the communication cables connected

to the instrument.

Step 5. Use a phillips head screw driver and remove the

two phillips head screws that fasten the back plate
to the rear of the instrument chassis. (See Fig. 3-14)

FIG. 3-14 REAR PLATE FASTENERS

Step 6. Grasp the two thumb screws and gently pull the rear

plate and Main Board out of the instrument.

Step 7. Use a phillips head screw driver and remove the

three phillips screws that fasten the rear plate to the
Single Channel Main board.

Step 8. Remove the two thumb screws that fasten the rear

plate to the Single Channel Main board.

Step 9. Attach the four (4) 7/8” standoffs to the Smart

Diagnostics Card with the four (4) 1/4” phillips
screws. Tighten the screws until they are snug. Do
not overtighten. (See Fig. 3-15)

Step 10. With the Hardy Logo facing the back of the Main

board, line up the pins of the Smart Diagnostics
card with the connector on the Main board.

NOTE:

Make sure that none of the pins are bent before
installing.

FIG. 3-15 STANDOFF LOCATIONS

Step 11. Gently push the pins into the connector making sure

that the standoffs are aligned with the standoff holes
in the Main board. When the standoffs are flush
with the Main board.

Step 12. Use a phillips head screw driver and install the four

(4) phillips screws that fasten the main board to the

Thumb
Screws

Phillips
Screws

SMART DIAGNOSTICS PWA
0535-0502-

REV

SN

OPTION

Install
Standoffs
Here

Install
Standoffs
Here

C o n n e c to r P in s

to M a in B o a r d