Test data display/tare limit - motion, Test data ingredient display/ingredient – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 134

Advertising
background image

HI-3010 Filler/Dispenser/IBC

118

Service Manual

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 25. Press the up or down arrow until the cursor is in

front of *Auto Zero Time.

The Auto Zero Time is the maximum
number of seconds that it takes the Auto
Zero to complete.

To change the Auto Zero Time press the
right or left arrow buttons to increase or
decrease the time.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 26. Press the down arrow button until the next three

menu items are displayed. (See Fig. 7-57)

FIG. 7-57 TEST DATA DISPLAY/TARE LIMIT -

MOTION TOLERANCE - AVERAGES

Step 27. *Tare Limit, *Motion Tolerance, *Averages can all

be changed from this menu.

Step 28. Press the up or down arrow until the cursor is in

front of *Tare Limit.

The Tare Limit is the maximum net weight
value that can be reduced (tared) to zero.

To clear the Tare Limit press the Clr
(Clear) button.

Use the alphanumeric key pad and enter
the new Tare Limit.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 29. Press the up or down arrow until the cursor is in

front of *Motion Tolerance.

The Motion Tolerance is the allowable
deviation between consecutive weight
readings, to confirm filling or dispensing
is active.

To change the Motion Tolerance press the
Clr (Clear) button.

Use the alphanumeric key pad and enter
the new Motion Tolerance.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 30. Press the up or down arrow until the cursor is in

front of *Averages.

The Averages is the Number of weight
readings used to compute displayed
weight.

To change the Averages press the Clr
(Clear) button.

Use the alphanumeric key pad and enter
the new Averages.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 31. Press the down arrow button until the next three

menu items are displayed. (See Fig. 7-58)

FIG. 7-58 TEST DATA INGREDIENT DISPLAY/

INGREDIENT - FILL CYCLES - TARGET WEIGHT

Step 32. *Ingredient, *Fill Cycles and *Target Weight can be

changed from this menu.

Step 33. Press the up or down arrow until the cursor is in

front of *Ingredient.

To select the ingredient press the right or
left buttons until the ingredient you want
to view or change appears.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 34. Press the up or down arrow until the cursor is in

front of *Fill Cycles.

The Fill Cycles are the number of fills to
complete one cycle.

Press the right or left buttons to increase
the number of fills per cycle.

Press the Enter button to set the entry.

Press the down arrow to move to the next
parameter you want to view or change.

Step 35. Press the up or down arrow until the cursor is in

front of *Target Weight.