Set clock parameter, About setting the clock, Setup menu/waversaver – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 58: Setup menu/set clock, Clock setup menu/set hours

Advertising
background image

HI-3010 Filler/Dispenser/IBC

42

Service Manual

effects of vibratory forces present in all industrial weight
control and measurement applications. By eliminating the
factor of vibratory forces the Filler/Dispenser is capable of

identifying the actual weight data. WAVERSAVER

®

enables

the Filler/Dispenser to distinguish between actual weight
data and mechanical noise, both of which are typically trans-
ferred to the Filler/Dispenser by the load cell signal.

WAVERSAVER

®

can be configured to ignore noise with fre-

quencies as low as 0.25 Hz. One of four higher additional cut
off frequencies may be selected to provide a faster instru-
ment response time. The function is user selectable and can
be turned off.

PARAMETER:WAVERSAVER

®

RANGE: .25 HZ, .50 HZ, 1.0 HZ, 3.50 HZ, 7.50 HZ,
OFF
DEFAULT: 1.00 HZ

Step 1. Press the Down arrow button until the cursor is in

front of WAVERSAVER

®

. (See Fig. 4-86)

FIG. 4-86 SETUP MENU/WAVERSAVER

Step 2. Press the right or left arrow buttons to select the set-

ting or turn WAVERSAVER off.

Step 3. Press the Enter button to set the entry.

Set Clock Parameter

About Setting the Clock

You set the Hour, Minutes, Month, Day and Year parameters
here. The settings are important when printing out the fill/
dispense data. These settings are the time stamps for the
alarms and fills/dispenses. The Timezone setting is used to
date the e-mails that are sent from the instrument in Green-
wich Mean Time (GMT).

PARAMETER:HOURS
RANGE: hh (01-24)
DEFAULT: NONE

PARAMETER:MINUTES
RANGE: mm (01-60)
DEFAULT: NONE

PARAMETER:MONTH
RANGE: Jan. thru Dec.
DEFAULT: NONE

PARAMETER:DAY
RANGE: dd (01-31)
DEFAULT: NONE

PARAMETER:YEAR
RANGE: yyyy (2001, 2002, 2003)
DEFAULT: NONE

PARAMETER:TIMEZONE
RANGE: -12 TO +12
DEFAULT: -8

Step 1. Press the Down arrow button until the cursor is in

front of Set Clock. (See Fig. 4-87)

FIG. 4-87 SETUP MENU/SET CLOCK

Step 2. Press the Enter button. The Clock Setup Menu

appears with the cursor in front of Set Hours. (See
Fig. 4-88)

FIG. 4-88 CLOCK SETUP MENU/SET HOURS

Step 3. Press the right or left arrow buttons to change the

Hours setting.

Step 4. Press the Enter button to set the entry.
Step 5. Press the Down arrow button until the cursor is in

front of Set Minutes.

Step 6. Press the right or left arrow buttons to change the

Minutes setting.

Step 7. Press the Enter button to set the entry.
Step 8. Press the Down arrow button until the cursor is in

front of Set Month. (See Fig. 4-89)

Advertising