Step 13. Press the Enter button. The Scale Weight for the

individual Scale Load Sensor display appears. (See
Fig. 7-108)

FIG. 7-108 INDIVIDUAL WEIGHT DISPLAY

If you do not get a reading from one of the
load sensors, either the cable to the load
sensor is disconnected or the load sensor is
malfunctioning.

If one of the load sensors is reading higher
or lower than the other load sensors in
your system and you know that you cali-
brated the instrument and cornered the
scale, there is something wrong with the
load sensor.

If one of the load sensors is reading higher
or lower than the other load sensors and
you do not know if the scale has recently
been calibrated, re-calibrate the scale then
check the Voltages and Weights again.

Stability Test

The Stability Test switches a fixed signal into the analog to
digital convertor, and calculates the mean squared variation
from the average reading, using 100 samples.
The test passes if the mean squared variation is less than 5.0,
and the average reading is between 30237 and 36955.

NOTE:

Before performing the Stability Test check to see
if all the Channels have been enabled from the
Web Browser.

Step 1. Press the up or down arrow buttons until the cursor

is in front of Stability Test. (See Fig.7-109)

FIG. 7-109 DIAGNOSTICS DISPLAY/SELECTING

STABILITY TEST

Step 2. Press the Enter button. The Information display

appears “SYSTEM STABILITY TEST, PRESS
ENTER TO START”. (See Fig. 7-110)

FIG. 7-110 STABILITY TEST DISPLAY

Step 3. Press the ENTER button to perform the stability

test. The display indicates Testing. (See Fig. 7-111)

FIG. 7-111 STABILITY TEST/TESTING DISPLAY

The results of the Stability Test are dis-
played for this scale as PASS. (See Fig. 7-
112) This means that the Mean Squared
Variation is less than 5.0 and the average
reading is between 30237 and 36955. In
short the instrument is working fine.

FIG. 7-112 SYSTEM STABILITY TEST DISPLAY/

PASS

The test results shown are for all four
Channels. If you want to see the results for
each individual channel, press the down or
up arrow buttons until the cursor is in front
of Test Results: PASS. Press the Enter but-
ton to see the test results per channel. (See
Fig. 7-113)

Channel 1 16.02 l b
Channel 2 16.03 l b
Channel 3 16.02 l b
Channel 4 16.00 l b

>

SMART DIAGNOSTICS
Vol t age & Weight - >
St abil it y Test - >
Fact or y Def aul t s - >

>

SYSTEM STABILITY TEST
Pr ess ENTER t o t est

!TESTING!
Pr ess Enter to Test

SYSTEM STABILITY TEST
Test Resul t s: PASS - >
Mean Sq. Var 2.37
Mean ADC Count 34k

>