Diagnostics, Voltage & weight displays, Load sensor display/serial number – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 137: Ensitivity - input resistance - output, Test data display/load sensor - ip address, Test and data menu

Advertising
background image

121

CHAPTER 7
Troubleshooting

Number of Sensors - The number of sen-
sors detected by the instrument in the Fill-
ing/Dispensing system.

1.

Load Sensor -The instrument will
read C2 load sensors certification
information only.

2.

The instrument displays sensor num-
ber 1 as a default.

3.

If you want to look at the certified
specifications for other load sensors
press the up or down arrows to move
the list of sensors. The instrument can
detect a maximum of 8 sensors.

4.

To view the certified sensor informa-
tion which is read from the C2 chip,
do the following:

a.

Press the Enter button. The Load
Sensor Display appears. (See Fig.
7-67)

FIG. 7-67 LOAD SENSOR DISPLAY/SERIAL NUM-

BER - CAPACITY - SENSITIVITY (MV/V)

b.

Serial Number - This is the serial
number of the selected load sen-
sor.

c.

Capacity - the maximum weigh-
ing capacity of the load sensor.

d.

Sensitivity - Sensitivity specifica-
tion of the load sensor set at the
factory.

e.

Press the down arrow until the
remainder of the load sensor
specifications appear. (See Fig. 7-
68)

FIG. 7-68 SENSITIVITY - INPUT RESISTANCE -

OUTPUT RESISTANCE

f.

Input Resistance - This is the
Certified Input Resistance from
the certification done by the fac-
tory.

g.

This is the Certified Output
Resistance from the certification
done by the factory.

Step 52. Press the down arrow button until the next three

menu items are displayed. (See Fig. 7-69)

FIG. 7-69 TEST DATA DISPLAY/LOAD SENSOR - IP

ADDRESS - DEVICENET ADR

Step 53. IP Address - Devicenet Address are read only.

IP - Lists the IP address for this instru-
ment. The address listed in Fig. 5-69 is the
default IP address of an instrument right
from the factory.

Devicenet Address - Lists the address of
the Node Address of the instrument you
are checking.

Step 54. Press the Exit button to return to the Test Data Dis-

play.

Step 55. Press the Exit button to return to the Test and Data

Menu. (See Fig.7-70)

FIG. 7-70 TEST AND DATA MENU

Diagnostics

Voltage & Weight Displays

Step 1. Press the down arrow until the cursor is in front of

Diagnostics. (See Fig. 7-71)