Operation - choose one/selecting, Operation - smart diagnostics, Peration - smart diagnostics/ weight – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 148

Advertising
background image

HI-3010 Filler/Dispenser/IBC

132

Service Manual

FIG. 7-117 HOME PAGE/SELECTING OPERATION

Step 2. Click on Diagnostics. (See Fig. 7-118) The Opera-

tion-Diagnostics Page appears with the four (4)
scales listed. (See Fig. 7-119)

FIG. 7-118 OPERATION - CHOOSE ONE/SELECT-

ING DIAGNOSTICS

FIG. 7-119 OPERATION - SMART DIAGNOSTICS

Step 3. Click on Weight and Voltage. (See Fig. 7-119) The

Operation/Diagnostics - Weight & Voltage page
appears with all four (4) channels (Inputs) with the
Sense Voltages, Millivolts, Millivolts/Volt, A/D
counts and Gross Weight displayed. (See Fig. 7-
120)

NOTE:

To differentiate between a unit with regular diag-
nostics and Smart Diagnostics (-SD) option card,
the regular diagnostics has only one input. Smart
Diagnostics has 4 inputs.

FIG. 7-120 OPERATION - SMART DIAGNOSTICS/

WEIGHT AND VOLTAGE PAGE

NOTE:

The values expressed here are for illustration
purposes only. Your values will be different.

NOTE:

The Weight and Voltage Web page shows all the
Weight, Voltage and A/D values at once. So if you
want to save time use the Web Browser, Smart
Diagnostics, Weight and Voltage Page for trou-
bleshooting.

The mV reading is a coarser reading than
the mV/V or Weight readings. The mV
reading is sufficient to balance the corners
of your scale or vessel.

These readings allow you to determine if
the problem is in the instrument (internal)
or in a load sensor(s) (external). The speci-
fication range for the Weight Controller is
0-15 mV. If you are getting a reading out-
side this range (15.5 mV, 3.1 mV/V Maxi-
mum or any negative values) the problem
is exterior to the Instrument (most likely
improper wiring). If you are getting a read-
ing between 0-15 mV the reading is nor-
mal.

Step 4. Check the results: