105 chapter 7 troubleshooting, F(a) – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 121

Advertising
background image

105

CHAPTER 7
Troubleshooting

F(a) - Verify Individual Load Cell Readings
Using Smart Diagnostics

Enter

Diagnostics

Voltage &

Weight

Activate the

test and

review the

results.

All signal

levels fall

within the +0

to +15mV

range

The Millivolt range is

outside the

+0 to +15 mV range

Check for

mechanical

reasons, or replace

load cell

Millivolt reading is

slow or 0.0 mV. It

may be a negative

reading. Use a multi-

meter to confirm

Replace the load

cell if the unit signal

readings are out of

tolerance.

B

Yes

F(a)

No

PASS

SMART DIAGNOSTICS is a built-in system diagnostics
utility that enables the operator to rapidly
troubleshoot a weighing system from the front panel,
PDA or Web Browser of the HI 3030. Used with an HI
SD card you can read each individual load cell in mV,
mV/V and weight to determine if a load sensor is
malfunctioning or not connected.

Apply weight
and insure all
signal voltages
increased the
same amount

Check
Wiring

Read and record

each load cell’s

Signal output

PASS

No