Waversaver® parameter – Hardy HI 6500 Series User Manual

Page 100

Advertising
background image

90 Chapter

5

On the HI 6500 series instrument display, select the Filter menu and use the UP or
DOWN button to select the NumAverages menu item. The current value will be displayed,
press the ENTER or the RIGHT button if this value needs to be modified.

WAVERSAVER® Parameter

Typically, mechanical noise (from other machinery in a plant environment) is present in
forces larger than the weight variations you want to detect. WAVERSAVER® reduces the
effects of the vibratory forces that exist in all industrial weight control and measurement
applications so the device can better calculate the actual weight. WAVERSAVER enables
the Weight Processor to distinguish between actual weight data and mechanical noise, both
of which are typically conveyed in the load cell signal. WAVERSAVER can be configured
to ignore noise with frequencies as low as 0.25 Hz. One of three higher additional cut off
frequencies may be selected to provide a faster instrument response time. The function is
user selectable and can be turned off.

RANGE: 0.25 Hz, 0.5 Hz, 1.0 Hz, 3.50 Hz, 7.50 Hz, OFF (default 1.0 Hz)

Chapter 1 describes the benefits of using the WAVERSAVER feature.

On the Web page, from the pull-down list, select the FILTER menu and then select the
value for WAVERSAVER.

On the HI 6500 series instrument display, select the FILTER menu and use the UP or
DOWN button to select WAVERSAVER menu item. The current value will be displayed.
If the WAVERSAVER selection needs to be modified, press the ENTER key, then use the
UP and DOWN buttons to select a new value and press the ENTER button to set and save
the new WAVERSAVER setting.

Advertising