Ref weight parameter – Hardy HI 6500 Series User Manual

Page 75

Advertising
background image

Instrument Configuration

65

Gravity Parameter

Gravity Correction compensates for an object weighing less at the equator than at the North
or South Pole.

Before modifying this value, please read the C2 calibration procedure described in Chapter
6.

On the Web page, select the calibration menu and left click inside the Gravity text field and
enter the desired Gravity value (see table in Chapter 6).

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the DO C2 eCAL submenu, then select the Gravity menu item. The
current value will be displayed, press the ENTER or the RIGHT button if this value needs
to be modified.

NumDevices Parameter

The NumDevices is a read only parameter that indicates the number of C2 eCAL load cells
that are connected to the HI 6500. This value is automatically calculated when Hardy C2
eCAL load cell(s) are detected during a C2 eCAL Calibration.

On the Web page, select the calibration menu and the number of C2 eCAL load cells
detected will be shown in the NumDevices text field.

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 eCAL submenu, use the IP or DOWN button the select the
NumDevices menu item and the number of C2 eCAL load cells detected will be shown on
the display.

Ref Weight Parameter

Reference Weight is the total live load that is currently on the scale. The calibration process
uses a reference weight, which is normally zero (no weight on the scale), but can be any
known weight on the scale. With nothing on the scale, the Reference Weight is 0.00. With
5 lbs on the scale, the Reference Weight is 5.00 lbs.

On the Web page, select the calibration menu and left click inside the Ref Weight text field
and enter the desired reference weight value.