Trouble shooting using integrated technician (it®), Stability test all, Trouble shooting using integrated technician (it – Hardy HI 6500 Series User Manual

Page 151

Advertising
background image

Troubleshooting 141

Trouble Shooting Using Integrated Technician (IT

®

)

The front panel display for the HI 6500 instrument’s Stability test is shown below. To
navigate this screen, select a test on the right using the right and left arrows with exit using
the left arrow when you are at the first out of the five positions.

The test displays are live readings and only available if you have an IT summing junction
card, maximum one J-boxes and 4 channels.

The test for pass or fail may take 30 seconds to settle. It will always show the load cells as
fail when the test is started.

The reduced voltage is a Yes or No selection. Then run the J-box test 1 as appropriate.

A more detailed explanation of each section of the test follows below:

Stability Test ALL

Stability: Gross = 0.9lb

RAW

Waversaver

Mean 32918

32993

Variation

0.17

0.00

Results

OK

OK

mV/V 0.0184
RTZ Pass

Advertising