Cal tolerance parameter – Hardy HI 6500 Series User Manual

Page 77

Advertising
background image

Instrument Configuration

67

On the Web page, enter the value in the text box provided.

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the Hard Cal submenu, then select the Cal Lo Weight menu item.
The current value will be displayed, press the ENTER or the RIGHT button if this value
needs to be modified.

Cal Tolerance Parameter

The value you enter for Calibration Tolerance sets the amount of deviation the instrument
will allow during the calibration process. This value must be greater than or equal to the
base motion value and/or the Graduation Sizes.

The base motion value can be calculated by using the following formula:

Base Motion value = (Total Load Cell Capacity x 0.0003)

We recommend three graduation sizes.

RANGE: 000001 - 999999 (default 10)

On the configuration Web page, enter the value in the text box provided.

On the HI 6500 series instrument display, select the calibration menu and use the UP or

DOWN button to select the Hard Cal submenu, then select the Cal Tolerance (Calibration
Tolerance) menu item. The current value will be displayed, press the ENTER or the RIGHT
button if this value needs to be modified.