Instrument Configuration

95

On the Web page, the value will appear on the Operations web page.

On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero Amount menu
item. The current value will be displayed.

Zero Command

When the Zero command is selected the Gross weight value is set equal to 0 (zero) if the
Gross weight is within the Zero tolerance, and the displayed weight goes to 0.00 if the
Gross weight is being displayed.

On the Web page, select the Operations menu and left click on the Zero button

On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero menu item and
press the ENTER button to execute this command.

NOTE

The Zero Command will fail if the scale is in motion

Auto Mode

Auto Mode enables one-button functionality when performing Zero and Tare operations,
automatically switching to Net mode when Tare is pressed. The default is Disabled.

On the Web page, select the Operations menu and left click to pull down the

Enabled/Disabled command to select Auto mode. Next select display mode Gross or Net.
Then click on the Save Parameters button.

On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to Enable or Disable Automode. For display mode, select Gross or Net.
Press the ENTER button to execute this command.

Count Operations Parameters and Commands

The count mode allows the unit to be calibrated and used as a counter. This mode allows
the user to "count" the number of pieces placed on the scale.