102 Chapter

5

On the HI 6500 series instrument display, select the Security menu and use the UP or
DOWN button to select the Read Only submenu. Enter the password and then change the
current password press the UP or DOWN button and enter the current password then arrow
to "password" and enter and use the arrow keys to enter new password and press the
ENTER button to save and exit this submenu; and return to the top level security menu.

Setup Parameter Menu

The Setup Menu consists of the following parameters:

Capacity

Decimal Point

Grads

Instrument ID

Motion Tolerance

Operator ID

Units

Capacity Parameter

Scale Capacity is the scale's nominal operating capacity (the total weight capacity of the
scale system). If this value is exceeded by six graduations, dashes appear on the front
display. Communications to and from optional devices are not affected.

If a Capacity value is entered which conflicts with either the decimal point value or the
graduation value, the decimal point and/or graduation values are automatically modified to
match the programmed capacity. It is therefore recommended that the Capacity parameter
is entered first before setting the Decimal Point and Grad parameters.

RANGE: .000001 - 999999 (Default 999999).

On the Web page, enter the capacity in the text box provided

On the HI 6500 series instrument display, select the Setup menu and use the UP or
DOWN button to select the Capacity item. The current value will be displayed, press the
ENTER or the RIGHT button if this value needs to be modified.

Decimal Point Parameter

Use the Decimal Point Parameter to set the resolution you want for the instrument. Here
you set the location of the decimal point for the weight resolution. The higher the number,
the farther to the left the decimal point moves and the higher the resolution of the scale.
Note that setting more decimal points does affect the overall accuracy of the instrument.