Hardy HI 6500 Series User Manual

Page 123

Advertising
background image

Calibration

113

C2 and eCAL Calibration from the Web Page

From the Home Page, click Configuration to display the Configuration menu; then
click Calibration to open the Calibration page.

Step 1. Check to make sure the number of load cells displayed on the webpage

corresponds correctly to the inputs on the instrument.

Step 2. In the “Ref Weight” text field, enter the reference weight for your application.

Step 3. An empty scale will use a 0.00 reference weight setting.

Step 4. (Optional) If you want to enter the Gravity Correction Factor, see the table

above, then enter the Correction Factor number you selected from the table in
the Gravity Correction field

Step 5. Click the “Do C2 Calibration” button

Step 6. Wait a few seconds and the results will appear. If the Calibration was successful

a Cal completed OK appears and gross weight will read your reference value.

However, if no load points are connected to the HI 6500 series the following

message appears. Check the cable and connectors of the load point(s) and re-
calibrate.

Step 7. This message appears if no load points are connected to the HI 6500. Check the

cable and connectors of the load point(s) and re-calibrate.

Cal failed: A to D converter error

Advertising