64 Chapter

5

On the Web page, enter the value in the text box provided.

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 (eCAL) submenu, then select the Cal Tolerance (Calibration
Tolerance) menu item. The current value will be displayed, press the ENTER or the RIGHT
button if this value needs to be modified.

Do C2 (eCAL) Command

When this command is selected the C2 eCAL calibration process is run using the Cal
Tolerance, Gravity Correction, Ref Weight, and Sensitivity parameter settings.

NOTE

Ensure the Sensitivity, Gravity and Reference weight are all set prior to running C2 cal
command.

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 (eCAL) submenu, then select C2 Cal. The instrument will
perform the calculation.

On the Web page, left click on the Do C2 eCAL button