Backlight parameter, Power saving parameters, Auto enable parameter – Hardy HI 6500 Series User Manual

Page 95

Advertising
background image

Instrument Configuration

85

o Display Mode
o Discover (Command)
o Num Display
o Display Order

Backlight Parameter

The Backlight parameter sets the display panel brightness level in 16 equal steps shown as
a progress bar on the instrument.

On the HI 6500 series instrument display, select the display menu and use the UP or
DOWN button to select the Backlight submenu. The current Backlight level will be
displayed, press the ENTER button if the value needs to be modified, and use the LEFT or
RIGHT buttons to set the required backlight level. Press the ENTER button to set and save
the backlight value.

Power Saving Parameters

To conserve power the display can be automatically dimmed towards a blank screen. While
the weight value is not in motion for the programmed Wait Time, the text is dimmed in 2
steps towards the blank level for the display. If the weight value is in motion, or the ENTER
button is pressed on the display, the value returns immediately to the preset background and
foreground levels.

Auto Enable Parameter

The Auto Enable parameter is used to turn ON or OFF the power saving mode.

On the HI 6500 series instrument display, select the display menu and use the UP or
DOWN button to select the Power Saving submenu, then select the Auto Enable menu item.
The current state, ON or OFF, will be displayed. Press the ENTER button if the state needs
to be modified. When modifying this parameter, ENTER toggles the parameter between the
two states ON and OFF, and pressing the ENTER button sets the state for the Auto Enable
parameter.

Advertising