Operator id parameter, Unit (of measure) parameter, Operator id parameter unit (of measure) parameter – Hardy HI 6500 Series User Manual

Page 115

Advertising
background image

Instrument Configuration

105

The base motion number can be calculated by using the following formula:

Base Motion Number = (Total Load Cell Capacity x 0.0003)

RANGE: .000001 - 999999 (default 10)

On the Web page, enter the value in the text box provided.

On the HI 6500 series instrument display, select the Setup menu and use the UP or
DOWN button to select the Motion Tolerance item. The current value will be displayed,
press the ENTER or the RIGHT button if this value needs to be modified.

Operator ID Parameter

The Operator ID is the ID of the user who is going to operate the Weight Processor or
service the instrument. Select three letters or numbers or any combination of letters and
numbers that adequately identifies the user.

On the Web page, type a name that identifies the operator of this instrument in Operator
ID text field.

On the HI 6500 series instrument display, select the Setup menu and use the UP or
DOWN button to select the Operator ID menu item. The current Operator ID will be
displayed, press the ENTER or the RIGHT button if this identifier needs to be modified and
use the UP, DOWN, LEFT, and RIGHT buttons to input a new identifier, then press the
ENTER button to save and exit this menu.

Unit (of Measure) Parameter

The Unit (of measure) parameter sets the scale to either English or Metric units. The
Selections are:

Ounces

Pounds

Ton

Gram

Kilogram

Metric Ton

Advertising