Tare command, Zero operations commands and parameters, Zero tolerance parameter – Hardy HI 6500 Series User Manual

Page 104: Zero amount parameter, Zero tolerance parameter zero amount parameter

Advertising
background image

94 Chapter

5

Tare Command

When the Tare command is selected the Net weight value is set equal to 0 (zero), and the
displayed weight goes to 0.00 if the Net weight is being displayed. If the Gross weight is
being displayed the displayed value does not change.

On the Web page, left click on the Tare button located in the Operations page.

NOTE

The Tare operation will fail if the scale is in motion.

On the HI 6500 series instrument display, select the calibration menu and use the UP or
DOWN button to select the Tare Operations submenu, then select the Tare menu item and
press the ENTER button to execute this command.

Zero Operations Commands and Parameters

Zero Tolerance Parameter

The value you enter for Zero Tolerance sets the weight unit limit from zero the instrument
will accept as gross zero during the zeroing function (when you push the Zero button).

NOTE

The amount of weight zeroed off is cumulative. The zero command will fail if the current
gross weight plus any previously zeroed amount exceeds the zero tolerance.

RANGE: .000001 - 999999 . (default 10.0)

On the Web page, select the Operations menu and enter the value in the text box provided.
Our example, for illustration only, sets the Zero Tolerance to 10.0

On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero Tolerance menu
item. The current value will be displayed, press the ENTER or the RIGHT button if this
value needs to be modified.

Zero Amount Parameter

The Zero Amount parameter is READ ONLY and displays the amount that has been
"zeroed" from the scale.