ALTANA LCM IV User Manual

Page 102

Advertising
background image

103

Section 8

Troubleshooting

Error displayed

Definition

Resolution

An error occurred when uploading

the instrument data.

Restart the process or contact the

manufacturer or national agent.

An error occurred when reading from

the USB memory stick.

Restart the process or contact the

manufacturer or national agent.

An error occurred when writing to the

USB memory stick.

Restart the process or contact the

manufacturer or national agent.

Please check on the current update

file.

Error during update.

Check the USB memory stick.

Please contact Customer Service.

Error during update.

Contact the manufacturer or national

agent.

Please check network configuration.

See

section 5.2.6.4, page 35

.

Please check the connection.

See

section 5.2.6.4, page 35

.

Please close the cover.

Close the cuvette compartment cover.

Please insert the USB memory stick.

Insert a USB memory stick into a USB A

port on the instrument.

Please check the connection and

contact the administrator.

Network setup or FTP error

Refer to

section 5.2.6.4, page 35

or

contact the manufacturer or national

agent.

File for instrument update missing.

Error during update.

Check the USB memory stick.

File for instrument update is faulty.

Error during update.

Save the update file again and repeat

the procedure.

It’s recommended to execute a

Full System Check

Check of the air values failed

Switch the instrument off and then back

on again. If the system test is

unsuccessful, contact the manufacturer

or national agent.

Entry invalid!

Password incorrect

Forgotten your password?
Contact the manufacturer or national

agent

Absorbance > 3.5!

The measured absorbance exceeds

3.5

Dilute sample and measure again

Color = ***

Color value outside measurement

range.

Dilute the sample or select appropriate

color scale.

Error when calling up the local IP

address.

Network setup: DHCP client has no

connection to the DHCP server

Enter the IP address again.

Error during default gateway setup.

Network setup: default gateway cannot

be set for fixed IP address

Try to create the connection again.

Error during network drive setup!

Error during network setup

Check the settings.

Error during subnet mask setup.

Network setup: Subnet mask cannot be

set for fixed IP address

Enter the subnet mask again.

Error copying from USB memory

stick.

Error during update

Restart the process or contact the

manufacturer or national agent.

Error in FTP connection.

FTP error

Make sure that the instrument is

connected to the network.

Advertising
This manual is related to the following products: