ALTANA LCM IV User Manual

Page 6

Advertising
background image

7

Section 1

Specifications

These are subject to change without notice!

Performance specifications

LCS IV

LCM IV

Display mode

Color measurement, color difference

measurement, absorbance and

concentration

Color measurement

Color measurement

26 color ratios

5 color ratios

Colorimetric evaluation

All visual color ratios are calculated for standard light chart C and 2° standard

observers in accordance with DIN 5033. Colorimetric color values can be switched

to light type A, C, D65 and 2° or 10° standard observers.

Source lamp

Halogen lamp

Wavelength range

320–1100 nm

Wavelength Accuracy

± 1.5 nm (wavelength range 340–900 nm)

Wavelength reproducibility

” 0.1 nm

Wavelength resolution

1 nm

Wavelength calibration

Automatic

Wavelength range for color

measurement

380–720 nm steps of 10 nm

Scanning speed

• 8 nm/sec (in steps of 1 nm)

Spectral bandwidth

5 nm

Photometric measuring range

± 3 Abs (wavelength range 340–900 nm)

Photometric accuracy

5 mAbs at 0.0–0.5 Abs,
1 % at 0.50–2.0 Ext

Photometric linearity

< 0.5 % to 2 Abs
” 1 % at > 2 Abs with neutral glass at 546 nm

Stray light

< 0.1 % T at 340 nm with NaNO

2

Data log

3000 color measurements,

100 color reference values,

1000 photometric measurements,

20 wavelength scans, 20 time scans

400 color measurements

Physical and environmental specifications

Width

350 mm (13.78 in)

Height

151 mm (5.94 in)

Depth

255 mm (10.04 in)

Earth

4200 g (9.26 lb)

Ambient operating requirements

10–40 °C (50–104 °F), maximum 80 % relative humidity (without condensate

formation)

Ambient storage requirements

–40–60 °C (–40–140 °F), maximum 80 % relative humidity (without condensate

formation)

Additional technical data

Power connector via external power

supply

Input: 100–240 V/47–63 Hz

Output: 15 V/40 VA