55

Standard programs

5.5.1 Take a color measurement

1. Press Color Measurement.

2. Insert a cuvette/sample cell with distilled water to calibrate.

Note: Always carry out the calibration very carefully, as a faulty

calibration can cause inaccurate results to be obtained.

3. The calibration starts automatically once the instrument has

detected the cuvette.

The type of cuvette/sample cell used and the exact progress

of calibration is shown in a separate window.

4. After calibration, the cuvette size used is displayed in the top

right.

Note: After calibration, you can measure the cuvette with distilled

water again as a sample.

The displayed measured values should match the uncolored color

indices (i.e. Hazen = 0, Gardner = 0.0, CIE L*a*b* = 100.0, 0.0, 0.0,

etc.)

5. Insert the test cuvette.

The measurement starts automatically.

The result of the color calculation is displayed.

Note: The bar on the right next to the result shows the result relative

to the measurement range.

6. For the next measurement, remove the cuvette and insert the

next sample cuvette, or press Measure to measure the same

sample again.

5.5.1.1 Touch-sensitive areas in measurement mode

In measurement mode, there are touch-sensitive areas that give

you immediate access to various menu options.