Advanced Operations

7. Select Options for further editing.

Select

New to define a new standard.

Select

Edit to change the settings.

Select

Delete to delete the standard.

Only user-programmed standards can be deleted.
Select

Reminder to enter an automatic interval at which you

are prompted to take a standard measurement.

6.1.4.2

Perform AQA Standard Measurement

If you have set a reminder function, you will be prompted to take

the AQA standard measurement after a set interval has elapsed.

1. Select the required AQA standard.

The AQA button in the toolbar is displayed in yellow.

2. Take the standard measurement.

6.1.5 Instrument backup

Prior to the next service date, the "

Instrument Backup" offers

the opportunity to store all programs, measurement data, user

IDs, passwords, as well as all adjustable data on a USB stick.

1. Press Instrument Backup in the "System Check" menu.
2. Insert the USB stick (refer to

section 3.5, page 16

).

3. Press Store to start a Backup.

If the file was stored, the message "Instrument Backup is stored"

will be displayed.
4. Press OK to return to the "System Check" menu.