ALTANA LCM IV User Manual

Page 60

Advertising
background image

61

Standard programs

3. Press Auto.

4. Press Graph.

The color matching result is now displayed graphically.
The upper value shows the option used for the color matching

calculation which can be

Auto for matching in all 5

Pharm.Eur.Scales or

B, BY, Y, GY or R for matching only inside of

the selected color scale.
The lower entry shows the color matching result, e. g. brown 4

(B4) in this example.
The 'Lab* value shows the color difference of the measured

sample relative to the stated reference solution B4.
The 'E* value shows the three-dimensional color difference of

the measured sample relative to the stated reference solution B4.

5. Press Return to change the option for the color matching

function

.

6. Select Scale is displayed.

7. The option for color matching to a special scale can be

changed.

8. Graph offers a visualisation of the matching result in the

selected matching option.

9. Depending on the scale selection, the appropriate graph of

the color scale will be displayed.

When option

Auto is selected the a*, b* color locus of the

measured sample will be displayed in all 5 color scales of the

Pharm Eur.

10. Press the Print symbol to print the graph.

Advertising