Standard programs

Integral
The Integral applies to the whole wavelength range of the scan.
1. Press Integral: Off in the "Options" menu.
2. Highlight On to show the Integral.
3. To find the integral of other wavelength ranges, change the

wavelength range and scan again.

4. Press OK to confirm.
5. Press Return to return to the Scan mode.

Note: In measurement mode, the integral is displayed where the date is

otherwise displayed.

Note: For the next scan reading the setting for the Integral will be On.

Scale & Units
1.
Press Options > More.. > Scale & Units.
2. Highlight the required units (Abs or %T).
3. Highlight Auto or Manual scaling on the graph's y-axis.

Note: When selecting the manual scaling, the limit values y

min

and y

max

can be entered using the alphanumeric keypad. The graph is adjusted to

display only the values in the selected range. If automatic scaling is

selected, the instrument sets the limits automatically so that the total

range can be displayed.

4. Press OK.
5. Press Return to return to the Scan mode.

5.7.4.2 Perform a wavelength scan

After selecting all scan parameters, record a base line (initial zero

reading). If a scan parameter is changed, a new base line must

be recorded. When the baseline has been scanned, the

instrument is ready to scan one or more samples.

1. In the main menu, press the Wavelength Scan menu option.
2. Insert the zero solution cuvette into the cuvette compartment

and close the cuvette compartment cover.

Note: During a scan, the cuvette compartment must be covered with

the cuvette compartment cover. Therefore, no scans can be

performed with 11-mm round cells.