ALTANA LCM IV User Manual

Page 82

Advertising
background image

83

Standard programs

3. Press Zero.

During the scan of the base line, "Zero calibration" is shown

in the screen.

4. Insert the prepared analysis cuvette into the cuvette

compartment and close the cuvette compartment cover.

5. Press Read.

Under the graph, the display "Reading..." appears, and a

graph of the absorbance or transmittance values for the

scanned wavelengths is continuously displayed.

The Wavelength Scan is complete, if

the graph is shown full-size,

the scaling of the x-axis occurs automatically,

the Cursor functions in the vertical navigation bar are

highlighted.

a signal sounds

Navigation of the wavelength scan graph or a wavelength

scan analysis

5.7.4.3 Work with reference scans

There are two options to work with

Reference Scan:

First option:
1.
Press Reference: Off in the "Options" menu to select

another scan to display on the same screen with the current

scan. Highlight the required scan number and press

Highlight reference.

Note: Once a reference scan has been selected, the Reference:

Off key in the "Options" menu turns into Reference: On.

Note: Only scans that have the same wavelength range and step

can be displayed using the overlay option. This process can be

repeated until all matching scans are displayed.

Table 12 Navigating the wavelength scan

Cursor function/

Zoom function

Description

Curve Icon

(Choice of Cursor mode)

Choice of Cursor mode

Peak/Valley (cursor moves between minimum/maximum absorbance

values) or Cursor mode

Tracking (cursor moves over each data point of the scan).

Arrow keys

The arrow keys (right/left) are used to move the cursor (depending on the selected mode) to the

next data point. The data of the data point (wavelength/absorbance or transmittance value) are

highlighted on the x and y axes.
Note: Press any point on the curve to display the associated data.

Zoom Icon

This function is used to magnify the section of the curve in the vicinity of the cursor. The original

curve size can be restored by pressing the zoom icon again.