Standard programs

Absorbance Formula
1.
Press Absorbance Formula.
2. The formula selected in the top key determines the number

of wavelength and coefficient keys that will appear below.

3. Press the key to change the calculation formula.
4. Select a formula from the displayed list.
5. Press OK.

When a new formula is selected, the number of variables

below changes to match.

The following formulas are available:

K

1

A

1

+ K

2

A

2

K

1

A

1

+ K

2

A

2

+ K

3

A

3

K

1

A

1

+ K

2

A

2

+ K

3

A

3

+ K

4

A

4

K

1

A

1

/ K

2

A

2

(K

1

A

1

+ K

2

A

2

) / K

3

A

3

(K

1

A

1

+ K

2

A

2

) / (K

3

A

3

+ K

4

A

4

)

A

1

refers to the absorbance at wavelength 1

A

2

refers to the absorbance at wavelength 2, etc.

O

Wavelength

To input the wavelength setting. Use the alphanumeric keypad to enter the reading wavelengths.

The entered wavelength must be in the range from 320–1100 nm.

Send Data icon

To send data to a printer, computer or USB memory stick (Type A)

Concentration Factor

Multiplication factor for converting absorbance values into concentration values.

Concentration

Resolution

To select the number of decimal places.

Absorbance Formula

Calculation basis for evaluating samples

Recall measurement

data

Recall saved measurement data, wavelength scans or time courses, see

section 5.3, page 38

.

Instrument Setup mode

Basic data of the instrument, refer to

section 5.2, page 23

.

Table 8 Multi wavelength setup options(Continued)

Options

Description

Advertising