1. Tap on Instrument Update in the "System Checks" menu.
2. Plug the USB storage device into the USB port (USB A) of

the device, refer to

section 3.5, page 16

.

3. Press OK. The link is established automatically and the

software is updated.

4. Press OK to return to the "System Check" menu.

Note: After the instrument update, a prompt to restart the instrument will

be displayed.

6.1.3 Optical Checks

The Verification Kit (refer to

Section 9, page 107

) is designed for

periodic monitoring of scattered light, photometric accuracy and

the wavelength accuracy of the spectrophotometers.
When results exceed allowable tolerances (given in the quality

control certificate to the test record), contact the manufacturer.

1. Press Optical Checks in the "System Check" menu.
2. Press Nominal Values.

3. Press Edit.

An automatic menu guidance queries values (filters,

wavelength, nominal values and tolerances) given in the

quality control certificate, to the following specifications:

Stray light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.

5. Press Verification.
6. Insert cuvette adapter Z (

Figure 3, page 18

) into the cuvette

shaft (2).