Standard programs

If air bubbles are entrapped, remove them by heat, vacuum,

ultrasonic treatment or other suitable means.

Clean the outside of the cuvettes/sample cells thoroughly

before inserting them in the cell compartment.

5.5 Color measurement

Proper sample preparation is extremely important for accurate

color measurement. To make sure that an exact measurement is

taken, refer to the following sample preparation guidelines:

Always clean the glass cuvettes/sample cells immediately

after use.

Only use optically preferred samples for measurement. Make

sure the cuvettes/sample cells are clean and show no signs

of opaqueness.

Slowly add the liquid to the cuvette in order to prevent the

formation of air bubbles in the sample.

Chapter

5.4, Page 53

contains further information on

preparing and taking samples.

The color measurement mode is used to determine absolute

color values in the Hazen, Gardner, CIE L*a*b* or European

Pharmacopoeia color scales.
For each cuvette type (11 mm round cuvette and 10 and

50 mm square cuvettes), an independent calibration data record

is used.
It is possible to calibrate the instrument with one, two or three

types of cuvettes/sample cells and to use these different

cuvette/sample cell types in parallel.

To use the 10 mm square cuvette and 11 mm round cuvettes,

adapter Z must be inserted into the cuvette compartment (2). For

measurements with 50 mm square cuvettes, you must remove

the adapter.

N O T I C E

Before using disposable cuvettes/sample cells made of PS

(Polystyrene) or PMMA (Polymethyl methacrylate), be sure that the

cuvettes/sample cells will not be destroyed by samples, otherwise the

cell compartment can be damaged.

Advertising