Standard programs

5.6.2 Take a color difference measurement with stored reference values

1. Press Color Difference Measurement from the Main Menu.

2. Press Select Reference.

3. A list of available references is displayed.

4. Press the required reference.

5. Press Select Reference to confirm.

6. Insert the test cuvette.

Important note: For the color difference measurement, it is

absolutely necessary that the type of cuvette/sample cell used for

the sample measurement is identical to the type of

cuvette/sample cell that was used for the reference

measurement.

Example:
It is not possible to select a reference that was measured in a

10 mm cuvette, and then run a difference measurement with a

50 mm cuvette.

5.6.3 Add a reference to the reference list

1. Press Color Difference Measurement from the Main Menu.

2. Insert a reference cuvette/cell.

3. Press Options >More > Store Reference. The Reference

Data list is displayed.

4. Press a free reference number.

5. Press Store.