Standard programs

5.7.4 Wavelength Scan mode – recording of absorbance and transmission spectrums

In the Wavelength Scan mode, the absorbance of the light in a

solution over a defined wavelength spectrum is measured. While

the scan is taking place, the cell compartment cover must be

closed. Therefore, a scan of the round cell compartment is not

possible.
The reading results can be displayed as a curve, as percentage

transmittance (%T) or as Absorbance (Abs). The collected data

can be printed as a table or a curve.
The data is available for formatting changes. These include

automatic scaling and zoom functions. Maximum and minimum

values are determined and shown as a table.
The cursor can be moved to any point on the curve for the

purpose of reading off the absorbance or transmittance value and

the wavelength. The data for each data point can also be

displayed as a table.

Table 10 Navigating the time scan

Cursor function/

Zoom function

Description

Curve Icon

(Choice of Cursor Mode)

Delta mode: A second cursor is highlighted. The position of the fixed cursor was previously

defined in Cursor mode Single. Use the active cursor to select any point on the reading curve.

The difference to the fixed cursor is shown on the curve. The delta values are correspondingly

highlighted and displayed on the x and y axes.

The gradient of the curve and the correlation coefficient (r

2

) between the cursor points in the

Delta mode are shown under the curve.
Cursor mode, single: The cursor moves to each selected measuring point of the scan.

Arrow keys

The arrow keys (right/left) are used to move the cursor (depending on the selected mode) to the

next data point. The data of the data point (wavelength, absorbance or transmittance value) are

highlighted on the x and y axes.
Note: Press any point on the curve to display the associated data.

Zoom Icon

This function is used to magnify the section of the curve in the vicinity of the cursor. The original

curve size can be restored by pressing the zoom icon again.