58

Standard programs

The current result of the actual reading and all further

measurements will be displayed in the selected color scale.

Alternative
1.
Tap the touch-sensitive area Result .

2. The result is shown in the next color scale in the list.

5.5.1.4 Change the measuring range after a measurement

1. Tap the touch-sensitive areas for the upper or lower limits of

the barchart.

2. Enter the new limit and press OK to confirm.

Note: The changed measurement range limit is displayed in blue.

Note: It is only possible to change the limits inside the measuring range

of the color scale. It is not possible to expand the range, e. g. to Hazen

2000.

The current result and all further measurements will be displayed

with the new measuring range.

5.5.2 Determination of the iodine color value

DIN 6162 defines the iodine color value as mg iodine per 100 mL

potassium iodide solution. Color matching with the iodine color

scale determines the depth of color of clear liquids like solvents,

plasticisers, resins, oils and fatty acids, whose color is similar to

that of a solution of iodine and potassium iodide of the same

thickness.
For iodine color values of less than or approximately 1, the

determination of the Hazen value according to DIN-ISO 6271 is to

be preferred. Taking and preparing samples is described in

Chapter

5.4, Page 53

. The iodine color value can be determined

with sufficient accuracy using an 11 mm disposable glass round

cuvette.

5.5.3 Determination of the Hazen color value (Pt-Co or APHA method)

The Hazen color value (ISO 6271, also referred to as the "APHA

method" or Platinum-Cobalt Scale) is defined as mg platinum per

1 liter of solution. The Hazen parent solution comprises 1.246 g

potassium-hexachloroplatinate (IV) and 1.00 g cobalt(II) chloride,

which is dissolved in 100 mL hydrochloric acid and topped up to

1000 mL with distilled water.
The Hazen-color scale is used to evaluate the colors of nearly

water-clear products. It has narrower gradations in the light

yellow range than the Iodine Color Scale and extends to

water-clear color casts.

Note: Samples with Hazen values of between 50 and 1000 can be

measured with sufficient accuracy in 10 mm square and 11 mm round