78

Standard programs

5.7.3.2 Time course scan reading

After all scan parameters have been selected, the instrument

must then be zeroed. Afterward the sample can be analyzed.
1. Insert the zero solution cuvette into the cuvette compartment

and close the cuvette compartment cover.

2. Press Zero.

The blank value is displayed on the screen.

3. Insert the analysis cuvette into the cuvette compartment and

close the cuvette compartment cover.

4. Press Read.

The instrument will begin collecting Time Course data (kinetic

data).

Note: During the reading the Zero and Read keys change to Mark

and

Stop.

Mark: Press the Mark button to mark the next measurement

point visibly on the screen. This mark is not used by the

instrument, but is available for the user and may indicate a

significant event, such as the addition of a sample or other

reagent. The mark is also shown in the table.

Select

Stop to stop taking sample readings.

5.7.3.3 Analysis of time course data

After the Time Course data has been collected, the following

manipulations can be done on the graphic data:

The

Time Course Program is complete, if

the instruments beeps after completion of the overall time

the graph is shown full-size,

the scaling of the x-axis occurs automatically,

the Cursor functions in the vertical navigation bar are

highlighted.

Navigation of a time scan or a time scan analysis
After a time scan has been completed, the time and the

absorbance/transmittance data is displayed as a curve.
Where the cursor is positioned on the curve, the elapsed time up

to this point and the corresponding absorbance are highlighted.