ALTANA LCM IV User Manual

Page 88

Advertising
background image

89

Advanced Operations

7. Remove any present cuvettes from the cuvette

compartment.

8. Close the cuvette compartment cover.
9. Press Start.

10. Follow the instructions on the screen and insert the different

filters one after another in the sequence shown.

11. Press Next after inserting a filter.

After the last reading the results are displayed.

12. Press PC & Printer icon to send the data to a USB memory

stick, PC or to a printer.
The files will be stored automatically as CSV files (Comma

Separated Value). The file name will be formatted as

"Verification.csv".

6.1.4 AQA Analytical Quality Assurance

6.1.4.1

Standard Configuration

Analyzing an AQA standard solution enables you to check your

working procedure, photometer and accessories.
1. Under "System Check", select AQA,

or

press

AQA in the toolbar on the right of the screen.

A list of standard solutions opens.