Standard programs 5.1.3 main menu – ALTANA LCM IV User Manual

Page 21

Advertising
background image

22

Standard programs

5.1.3 Main menu

A variety of modes may be selected from the Main Menu. The

following table briefly describes each menu option.
There is a toolbar on the right-hand side of the screen. Press to

activate the various functions.

Table 3 Main Menu options

Option

Function

Color measurement

The

COLOR MEASUREMENT MODE

is used to determine color values such as Hazen,

Gardner and Saybolt. The LCS IV also offers three-dimensional, absolute colorimetric

values, as well as the color scales of CIE L*a*b*, Hunter Lab or the European

Pharmacopoeia.

Color difference measurement

(only LCS IV)

The mode for

COLOR DIFFERENCE MEASUREMENT

is used to determine a

quantitative color difference between a reference (R) and a sample (P) in the

three-dimensional color space (CIE L*a*b* or Hunter Lab). In this mode, an additional

reference memory for up to 100 references is available.

Photometry

(only

LCS IV)

Single Wavelength

Single wavelength readings are:
Absorbance readings: The light absorbed by the sample is measured in absorbance

units.
Transmittance reading (%): Measures the percent of the original light that passes

through the sample and reaches the detector.
Concentration readings: A concentration factor can be entered to enable the

measured absorbance values to be converted into concentration values.

Multi Wavelength

In the Multi Wavelength mode, absorbance (Abs) or percentage transmittance (%T) is

measured at up to four wavelengths and absorbance differences and absorbance

relationships are calculated. Simple conversions into concentrations can also be

carried out.

Time course

The time scan records the absorbance or % transmittance at a wavelength over a

defined time.

Wavelength Scan

A wavelength scan shows how the light from a sample is absorbed over a defined

wavelength spectrum. This function can be used to determine the wavelength at

which the maximum absorbance value can be measured. The absorbance behavior is

displayed graphically during the scan.

System Checks

The "System Check" menu contains a number of options, such as instrument

information, optical checks, instrument backup, service times, instrument update,

settings for analytical quality assurance and lamp history.

Recall measurement data

Saved data can be retrieved, filtered, sent to a printer, memory stick or PC and

deleted.

Instrument Setup

This menu is used to configure user-specific and/or process-specific settings:

operator ID, date and time, security settings, saved data, sound, PC and printer and

energy management.

Advertising
This manual is related to the following products: