Standard programs

4. Press Options and then the PC & Printer icon.

5. Select the data to be sent.

The following options are available:

Single point: only this selected reading will be sent

Filtered data: only readings that correspond to the set

filters will be sent

All data: all data in the selected data category will be

sent.

6. Press OK to confirm.

Note: The number in parentheses is the total number of data sets

assigned to this selection.

5.3.3.4 Delete single and multi-wavelengths from the measured data memory

1. In the main menu, press the Recall Data menu option.

2. Press Data Log > Options > Delete.

3. Highlight Single point, Filtered data or All data and press

OK to confirm.

Note: The number in parentheses is the total number of data sets

assigned to this selection.

5.3.4 Wavelength scan and time course (LCS IV only)

You can save 20 data records for the wavelength scan and 20

data records for the time course. The data must be stored

manually after it is displayed.