4 take and prepare samples – ALTANA LCM IV User Manual

Page 52

Advertising
background image

53

Standard programs

9. Press Add Data to include additional measured values in the

trend.

5.4 Take and prepare samples

Take a representative sample from the product you wish to

measure in accordance with DIN EN ISO 15528 (or ASTM

D3925-02).
If the material shows any signs of turbidity, remove this turbidity

by filtration, centrifuging, heating, ultrasound treatment or other

suitable means.
Heat partly solid samples before measuring in order to dissolve

the solid material in the liquid. The preparation must not cause

any chemical changes in the sample.
Make sure that during the measurement there are no bubbles in

the sample.
For the color difference measurement, three cuvette types are

available that differ in terms of material (glass, PS and PMMA)

and path length (10 mm, 11 mm and 50 mm). Add approx. 2 cm

of the sample to the cuvette. The light beam passes through the

cuvette approx. 0.5 cm to 1.5 cm above the bottom of the cuvette.
The program calculates the iodine, Hazen, Gardner, Saybolt,

Klett and ASTM D 1500 color values automatically and displays

the color values. The cuvette type being used is taken into

consideration.
A dry thermostat is available for the round disposable glass

cuvettes with 11 mm. The dry thermostat heats up the cuvette to

any temperature between ambient temperature and 150 °C

(302 °F).

Always hold the cuvette/sample cell close to the top to make

sure that there are no fingerprints in the measurement zone

of the cuvette/sample cell. Use suitable transfer pipettes to

introduce samples into the cuvettes/sample cells.

Slowly add samples to the cuvettes/sample cells to make

sure air bubbles do not form on the cuvette/sample cell wall

and in the sample. Air bubbles will cause false readings.

N O T I C E

The samples must be clear and free of turbidity. If products in paste or

solid form cannot be measured directly, the product must be melted

before being transferred to the cuvettes/sample cells. Make sure the

cuvettes/sample cells do not contain any air bubbles.

Advertising
This manual is related to the following products: