Troubleshooting – ALTANA LCM IV User Manual

Page 103

Advertising
background image

104

Troubleshooting

Fault

Test program stopped!

Please check lamp

Close the lid.

Error [xx]

Test program stops when the instrument

is started

Check the lamp and replace it if

necessary.
Close lid.
Press

Start Again.

Fault

Test program stopped!

Please remove the cuvette

Close the lid.

Test program stops when the instrument

is started

Remove the cuvette/sample cell from

the cell compartment.
Press

OK.

Error

Selfcheck stopped.

Hardware error.

Error [x]

Electronic defect

Contact the manufacturer or national

agent and specify the error number.

Error

Too much ambient light!

Move instrument into shade

or close the lid

The instrument sensors detects too

much ambient light.

Decrease ambient light. Avoid direct

sunlight.
Close lid.

No instrument backup present!

Check the USB memory stick.

No valid data for these parameters!

Data analysis not possible, no

measurement data

Change the selection.

No valid data found!

Data cannot be viewed in the data log

Change the selection.

No measurement data present!

Data analysis settings cannot be

configured without measurement data.

Change the selection.

Control range not reached!

Data analysis limits not reached

This is a warning notice. The control

limit set was not reached.

Control range exceeded!

Data analysis limits exceeded.

This is a warning notice. The control

limit was exceeded.

Concentration too high!

Calculated concentration is higher than

999999

Dilute sample and measure again

Possible interference by:

Interference Check

The analysis is possibly erroneous due

to interferences.

Possible interference from:

Interference Check

The analysis is possibly erroneous due

to interferences.

Next service is due!

Contact the manufacturer or national

agent for an instrument service. For

setting the service reminder, refer to

section 6.1.7, page 92

Negative result!

The calculated result is negative

Check concentration of sample

Network switched off.

Network setup off, when access to

homepage via sidebar

Activate the online connection.

Remote server cannot be reached.

Error during network setup

Make sure that the instrument is

connected to the network.

Unstable lighting conditions!

Avoid direct sunlight at the measuring

location.

Please insert adapter Z.

For measurements with 11 mm round

cuvettes, adapter Z is required.

Insert adapter Z into cell compartment

#2. Confirm with OK.

Insufficient memory for update .

Error during update.

Select a memory with more space.

Error displayed

Definition

Resolution