Menu maps – YSI 5200A User Manual

Page 109

Advertising
background image

YSI 5200A

217

YSI 5200A

216

Menu Maps

continued next page

Run Screen

 → 

Menus

 → 

Sensor Setup

Conductivity

Conductivity (Enabled)

Conductivity (Enabled)

Control Mode (Set Point)

T

Set Point

                        PID/PWM

Alarms (Disabled)

Alarm Low (460uS) 1 to 10000*



Alarm Low Output (disabled)

Alarm  Low (Disabled)

Instrument (Local Machine)
Relay (Relay 1)

T

Relay 1

                Relay 2
                Relay 3
                Relay 4

Set Point control only

Range Low (480 uS)
461 to 10000*



Range Low (Disabled)

Instrument (Local Machine)
Relay (Relay 1)

T

Relay 1

                Relay 2

Relay 3

               Relay 3

                Relay 4

Dose Up (30 Sec)
1‐255



     Time Units

T

Sec

                        Min
                        Hours

Dose Up Wait (15 Min)
1‐255



     Time Units

T

Sec

                        Min
                        Hours

Set Point (500 uS)
481 to 10000*



PID/PWM control only

Control Up/Down (Rising)

T

 Rising

         Falling

Menu Maps

Run Screen

 → 

Menus

 → 

Sensor Setup

Conductivity continued from previous page

PID/PWM control only

P.I.D. Interval (100.00)
1.00 to 999.90



PID/PWM control only

P.I.D. Reset Interval (20.00)
1.00 to 999.90



PID/PWM control only

P.I.D. Gain (2000)
0.00 to 10.00
1.00 to 999.90



PID/PWM control only

P.I.D. Relay (Disabled)

Range Low (Disabled)

Instrument (Local Machine)
Relay (Relay 1)

T

 Relay1

         Relay2
         Relay 3
         Relay 4

Set Point control only

Range High (520uS)
501 to 10000*



Range High Output (disabled)

Instrument (Local Machine)
Relay (Relay 1)

T

Relay 1

                Relay 2
                Relay 3
                Relay 4

Alarm High (540uS)
521 to 10000*





Alarm High Output (disabled)

Alarm  High Output enable/disable

Instrument (Local Machine)
Relay (Relay 1)

T

Relay 1

                Relay 2
                Relay 3
                Relay 4

Set Point control only

Dose Down (30 Sec)
1‐255



     Time Units

T

Sec

                        Min
                        Hours

Dose Down Wait (15 Min)
1‐255

     Time Units



T

Sec

                        Min
                        Hours

Temp Comp % (1.91%)
0.00 to 4.0