Configuring the 5200A

6. Press

for feed timer relay to energize for configured run test time.

Countdown of relay “on” time is displayed.

7. When run time expires, the relay de-energizes. Enter the amount of food

dispensed at the Measured Weight screen.

8. Press

to save settings and return to Run Screen.

Notes:

- Since feeders and food size vary, run several calibrations and enter the

average weight at the last calibration. If feeder or feed type is changed,

recalibrate the feeder.

- If no test/calibration has been performed or if the results have not been

saved, the 5200A will be unable to perform any feeding(s) because there

will be no reference to determine the rate at which feeding(s) will oc-

cur.

- Stored calibration data is available from the AquaManager program.

- If feed timer units change, the feeder must be recalibrated.

- If an incorrect value is entered and saved, another calibration must be

performed.

- Disable the feed timer system when performing the feed timer calibration,

de-energizing any active feed timer relay.

- After calibration is complete, press

to Run Screen. The feed timer

system resets based on the new calibration configuration.

- Feed Timer calibration cannot be done using AquaManager.

AquaManager and the Timer Menu

Most Timer menus can be configured using AquaManager. After polling the 5200A

node, use “Systems Management” to view and upload 5200A configurations. Select

the Timers/Feeder tabs to view and change configuration.

Configuring the 5200A