Data logging, Event logging – YSI 5200A User Manual

Page 54

Advertising
background image

YSI 5200A

107

YSI 5200A

106

Notes:

- Use AquaManager to store and analyze 5200A logs. To avoid gaps in data

use polling intervals based on the user defined data logging configuration.

Set the data log interval so that analyzing the data is meaningful.

- AquaManager users can view Configuration events in the

Configuration log. Configuration

events are created when certain 5200A

configuration changes are made. The configuration log is not accessible

at the 5200A. See AquaManager Online Help for additional information.

Data Logging

Menus → System →Data Logging→Data Logging

The 5200A sensor log stores records for sensors and aux systems. Sensor logs are

time and date stamped based on the user configured logging interval. The logging

interval is user configured at 1-255 seconds, minutes or hours.

1. Enter Data Logging menu

Menus → System →Data Logging→Data Logging

2.

Use

and

to scroll and highlight submenus.

3.

Press

to select.

4. Enable and configure submenus.
5. Press

to save configuration and return to Run Screen.

Press

to view the 25 most recent sensor logs.

Notes:

- Temperature values are stored in °C regardless of display format.

- Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged

as Aux 1.00.

- The 5200A sensor log records 0.00 resolution for aux analog inputs regard-

less of the selected display format.

- DO values are stored in mg/L & %Sat regardless of selected display

format.

Configuring the 5200A

- Sensor logs record the default system name and not the user configured

system name.

- Changes to the data logging interval take effect after the expiration of the

interval being changed. Example - if the interval is changed from one

hour to one minute, one minute logging will begin after the expiration of

the current one hour interval. Power cycle the 5200A to make a change

to the data logging interval occur immediately.

- Menu configurations are saved to the system by pressing

to

return to the Run Screen and not when the change is made in the menu.

Exception - Date and time and clear data logs are updated immediately
to the system, not when

is pressed.

- The sensor log contains values for all sensor and aux systems regardless

if the system is disabled.

- 5200A sensor log values for disabled systems and systems without sen-

sors are not valid.

- AquaManager data log displays 5200A conductivity values in µS to 0.00

resolution.

In the above sensor log screen shot ORP is disabled.

Event Logging

Menus → System → Data Logging → Event Logging

Control, Alarm, and Timer relays assigned to 5200A systems create events when

they turn on and off. In addition, AC power failures, ethernet device error messages,

and slave network failures are create events. These events are stored in the Event

Log. Event logs identify the 5200A system, are time and date stamped and record

relay the action (“on” or “off”).

1. Enter Event Logging menu

Menus → System → Data Logging → Event Logging

2.

Use

and

to scroll and highlight submenus.