- All aux inputs have input protection, the maximum input voltage should

not exceed 6 VDC.

- Analog systems will not monitor and/or alarm properly if the device that is

installed and configured operates outside the configured analog range.

- For proper digital input operation, use low impedance switches and wir-

ing. The total resistance should not exceed 200 Ω (ohms).

Directions to Wire Aux Inputs

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant

d’effectuer un câblage quelconque.

1. Perform steps 1 - 4 of 5200A installation.

2. Open front panel - page 39 .

3. Feed aux device wires through drilled hole in rubber grommet to locations

K.

Terminal strip is removable.

CAUTION:

Run high and low voltage cables through separate bulkhead and

conduit.

4. Use (1) or (2) pin to terminate one wire and (G) to terminate ground wire.

(G) pin is the common ground for Aux 1 and 2.

CAUTION

: Aux inputs are not isolated. Devices connected must be electrically

isolated from ground and the water.

Installation and Wiring

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.21

5. Complete other wiring to 5200A.

6. Close front panel - page 40.

7. Test aux systems.

Test Auxiliary Input Devices

Apply power to 5200A. See Aux Setup - beginning on page 51 for configuration

information. Verify display, control, and/or alarm functionality based on user-

defined configuration.