YSI 5200A

143

YSI 5200A

142

Reconfiguration of the following does not place the system in 30 second hold off:

- DO systems - display format, membrane type, manual ppt input and

elevation.

- Temperature system - display format, chiller on time, chiller on time

units, chiller hold off time and chiller hold off time units.

Notes:

- Configuration changes place only the system that has been reconfigured

on 30 second hold.

- No “off” event log is created when energized system relays de-energize

due to 30 second sensor hold off.

- Control icons and control messages continue to display during the 30 sec-

ond sensor system hold at the Run Screen regardless if relays are energized.

Aux Hold Off

Aux Hold Off (user configured) is the length of time the control or alarm condition

must exist before the Aux relay energizes.

Notes:

- To avoid control or alarm for spurious events , configure the Alarm Hold

Off > than 0.

- If General Alarm is enabled, it is triggered after the Aux hold off time and

not the general alarm hold off time.

Aux Digital

Configure digital control/alarm (open/closed), relay, alarms, and hold off in the Aux

1 - Aux 2 Sensor Setup Menu. Use digital input type for control, or alarm, based on

either a switch open or switch closed condition. Only one relay can be configured in

an aux digital system. Enabled relay will energize based on the condition configured

in the control/alarm on menu.

Aux Relay Digital

Menus → Sensor Setup→Aux 1-2

When enabled, Aux digital relay energizes when configured control/alarm on condi-

tion occurs. Aux relay energizes after configured aux hold off time. When alarms

are disabled and as long as control/alarm on condition exists, the relay remains

energized including when menus are entered. Relay de-energizes when the non

control/alarm condition is processed by the 5200A regardless if alarms are enabled.

If alarms are enabled, the relay also de-energizes when one of the four front panel

Configuring the 5200A

softkeys

is pressed. Pressing one of four softkeys

resets the alarm system. The Alarm system becomes active again if, after the aux

hold off period, the alarm condition still exists.

Aux digital events (event log) are identified by aux number, input type (digital), relay

condition (on/off), and do not identify digital condition (open, or closed). Aux digital

events are not identified as control or alarm events regardless if alarms are enabled.

Directions to Configure Aux (digital)

1. Wire Aux digital devices - page 49.

2. Configure “digital” at Menus → System → Configure Aux →Input Type

3. Enter Aux 1-2 in Sensor Setup menu

Menus → System → Sensor Setup → Aux 1-2

4.

Use

to scroll and highlight submenus.

5.

Press

to select.

6. Enable and configure submenus.
7. Press

to save configuration and return to Run Screen.

Notes:

- It is important to keep the 5200A at the Run Screen during normal

operation. Alarm functionality (email alarms, activation of enabled

alarm relays, alarm icons, and buzzer) is suspended when 5X00 is not

at Run Screen.

- At sensor log Aux (1:1.00 ) is reporting open condition. Aux (1: 0.00) is

reporting closed condition.

- An aux digital relay will energize when alarms are enabled if in menus

longer than the aux alarm hold off time and the control/alarm condition

is processed from an existing non control/alarm condition. The general

alarm will not trigger until the user exits to run screen, the control/alarm

condition is reset by processing a non control/alarm condition, AND then

a control/alarm condition is reproduced.

Configuring the 5200A