Reset to factory – YSI 5200A User Manual

Page 63

Advertising
background image

YSI 5200A

125

YSI 5200A

124

Reset Clean Probe timer at the

menu or the Reset Clean Probes menu.

When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the

user configured number of days.

Reset To Factory

Menus → System → Service→ Reset To Factory

The Reset to Factory menu resets some, or all, of the 5200A systems to factory

defaults. For example, if the Clean Probe Timer is user configured to 60 days,

performing a factory reset will reset the Clean Probe Timer to 90 days which is the

factory default setting and not the user configured value. See Appendix B Menu

Maps - starting on page 201 for factory default settings for all systems.

Configuring the 5200A

Notes:

- “Reset All” and “Reset All Systems” resets time format to 12 hour and

date format to MMDDYYYY.

- It is strongly recommended that user configuration be maintained in a

log by the user.

- Instrument, aux, sensor, and suffix names are not reset at any reset to

factory menus. Relay names are reset at “reset relays” and “reset all”

menus.

- Perform sensor calibration after selecting “Reset All” and/or “Reset Sen-

sors”.

- Control icons and control menu messages are displayed for 30 seconds

after pressing

from reset to factory menus for any system that

was active (energized relay) prior to the factory reset.

Reset All

Menus → System → Service→ Reset To Factory→ Reset All

Resets all systems, sensors, and timers to factory defaults.

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. When prompted, “Are you sure?”, select one of the following to confirm,

4. Press

to save configuration and return to Run Screen.

Notes:

- Display contrast, backlight timer, and buzzer can only be reset to factory

defaults through this menu.

- Sensors must be recalibrated after performing a factory Reset All and/

or Reset Sensor.

Reset System

Menus → System → Service→ Reset To Factory→ Reset System

Specific systems or all systems listed in the Reset System menu can be reset to fac-

tory defaults.

Configuring the 5200A

Advertising