Input system specifications – YSI 5200A User Manual

Page 8

Advertising
background image

YSI 5200A

15

YSI 5200A

14

Input System Specifications

Aux Systems

There are 2 Aux input systems that allow digital or analog (0-

1.0VDC, 0-5.0VDC, or 4-20mA) configuration.

Aux System

Type

Digital

Analog

Range

dry contact open/

close

0 -1.0 VDC, 0-5.0 VDC, or 4-20mA

Accuracy

n/a

0-1.0VDC: ± 3% ± 10 mV

0-5.0VDC: ± 3% ±15 mV

4-20 mA: ± 3% ±100μA

Resolution

n/a

user selectable (0), (0 .0), or (0.00)

Probe assembly 5562 (DO, temperature, pH, ORP, and conductivity sensors)

Probe assembly 5561 (DO and temperature sensors)
Cable grade

military-spec (MS) connector

Cable lengths

4, 10, 20 m (special order up to 100m for DO and temp only)

Temperature
Sensor type

10,000 ohm @ 25°C thermistor

Range

0-45°C (32-113°F)

Accuracy

± 0.2°C

Resolution

0.1°C

Dissolved Oxygen
Sensor type

Membrane covered, steady-state polarographic

Range

0 to 500% air saturation

0 to 60 mg/L

Accuracy

±2% or ± 0.2 mg/L, whichever is greater for 0-200%

±6% of reading for 200-500%

Resolution

0.1% (0.0-99.9%); 1% (> 99.9)

0.01 mg/L (0-10 mg/L); 0.1 mg/L (10-60 mg/L)

Response Time

95% of reading in 17 seconds for 2.0 mil PE membrane;

95% of reading in 36 seconds for 2.0 mil Teflon membrane.

Allow approximately 10-15 minutes after 5200A is powered

on, or the DO sensor is serviced, for the polarographic sensor

to stabilize.

Introduction

Flow requirements minimum flow rate of 3 inches per second or 7.62 cm/second

for 2.0 mil PE membrane; 6 inches per second or 15.24 cm/

second for 2.0 mil Teflon membrane

Other

Automatic salinity compensation for DO readings when a

conductivity sensor is in use and enabled. Manual salinity

compensation for DO readings when a conductivity sensor is

not in use.

pH
Range

0.00 to 14.00 pH

Accuracy

± 0.1 pH

Resolution

0.01 pH

ORP
Range

-2000 mV to +2000 mV

Accuracy

± 20 mV

Resolution

1 mV

Conductivity
Sensor type

4 electrode, Cell constant of 5.0 ±5%

Ranges*

0 to 200 μS

0 to 2000 μS

0 to 20 mS

0 to 200 mS

Accuracy

0 μS to 100 mS, ±0.5% of range

100 mS to 200 mS, ± 1.0% of reading

Resolution

range dependent

0.1 μS (0 to 200 μS)

1.0

μS (0 to 2000 μS)

0.01 mS (0 to 20 mS)

0.1 mS (0 to 200 mS)

Salinity
Parameter Source

1) calculated from temperature & conductivity systems when

enabled

OR

2) user ppt input in DO sensor setup menu

Range

0-80 ppt

Note: When not using a conductivity sensor, the manual

input range is 0-50 ppt.