Updating 5200a firmware, Firmware updater directions – YSI 5200A User Manual

Page 57

Advertising
background image

YSI 5200A

113

YSI 5200A

112

1. Enter Display Settings menu

Menus → System → Display Settings →Display Format

2.

Use

and

to scroll and highlight submenus.

3.

Press

to select.

4. Press

to save configuration and return to Run Screen.

Notes:

- At 1x2 & 2x2 display settings, analog aux values over 7 digits will overlap

on to the adjoining sensor window. Adjust aux display format to lower

resolution.

- Because of font limitations, 1X2 and 2X3 display formats do not show

all 20 characters of a 20 character name. Typically only 13 characters

are displayed.

Updating 5200A Firmware

Like all microprocessor based instruments, it may be necessary to upload revisions

to the 5200A firmware. Download new firmware using the Firmware Updater utility

in AquaManager. A serial RS232 connection is necessary to run the Downloader

program.

The instrument firmware may become corrupted if any of the following occurs

during a 5200A firmware update:

- Firmware Updater utility is closed;

- AquaManager is closed;

- RS232 cable is removed from 5200A or PC com port;

- power is disconnected from 5200A.

If the 5200A’s firmware becomes corrupted and is not updated properly, the instru-

ment will need to be returned to YSI

.

Configuring the 5200A

Firmware Updater Directions

1. Wire RS232 connection. See RS232 Communication - page 59. (Dis-

regard if already using RS232 connection.)

Note:

Connect RS232 directly into slaves in order to upload firmware to slaves.

Disable slaves at the master to avoid slave failure alarms when uploading

- Menus → System → Communications → Network →Master.

2. Enable the Force Serial Port on 5200As wired and configured with an

ethernet device - page 100.

3. Verify that the baud rate in 5200A Communications menu is configured

to 115200.

4. From the AquaManager Explorer window, access the Firmware Updater

utility from the pull down menu: Tools→ Instrument tools→Update

Firmware.

5. Configure PC com port number (1-8) in Flash Programmer.

Notes:

- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS

YSI p/n

773

.

- When using USB-Serial Adapter, verify that the port number assigned

during driver installation is the same port configured when using Aqua-

Manager Flash Programmer.

Configuring the 5200A