Two range controls for do, Directions to configure do sensor, Directions to configure orp sensor – YSI 5200A User Manual

Page 75

Advertising
background image

YSI 5200A

149

YSI 5200A

148

- 5561 probe/cable users should ALWAYS disable the conductivity system

since there is no conductivity sensor in use to accurately compute a

salinity value.

- Changes to salinity affect the DO mg/L value.

- Changes to salinity do not affect the DO %Sat value.

- The DO system uses the temperature sensor value regardless if tempera-

ture is enabled. It is important to verify that the temperature sensor is

measuring accurately.

Two Range Controls for DO

The 5200A DO system has two low and two high control ranges. This dual range

allows for more control over oxygen levels in the system stream. Example - DO

range low system becomes active, an air injection system energizes when properly

wired and configured to relay #1. The DO level continues to fall. Range low two

system becomes active and a pure oxygen injection system energizes when properly

wired and configured to relay #2.

Directions to Configure DO Sensor

1. Connect probe assembly - page 61.

2. Enter DO Sensor Setup menu

Menus →Sensor Setup→DO

3.

Use

and

to scroll and highlight submenus.

Configuring the 5200A

4.

Press

to select.

5. Configure submenus.
6. Press

to save configuration and return to Run Screen.

Notes:

- DO set point, control, and alarm values are always entered in mg/L re-

gardless of display format.

- DO set point, control, and alarm values can be up to three (3) characters,

excluding decimal point. Values will truncate to 10ths when 10s digit

is configured.

- 5200A calculates the correct %Sat value based on configured elevation

during calibration.

- When DO range low 2 or DO range high 2 are active with energized

relays, only the range low 2 and range high 2 control messages and icons

display at the control menu and Run Screen regardless if DO range low

1 or DO range high 1 relays are also energized.

- See 30 Second Sensor Hold Off - page 141.

ORP

Menus →Sensor Setup→ORP

Read Sensor Setup Control and Alarms - page 132 before configuring ORP system.

Directions to Configure ORP sensor

1. Connect probe assembly - page 61.

2. Enter ORP Sensor Setup menu

Menus →Sensor Setup→ORP

3.

Use

and

to scroll and highlight submenus.

4.

Press

to select.

5. Configure submenus.
6. Press

to save configuration and return to Run Screen.

Configuring the 5200A