Set Point Control

Set point control menu options are shown below for an aux analog system.

Control relays configured in the Set Point control mode energize when a user config-

ured control value is received from the input device and is processed by the 5200A.

Control relays can be configured to raise and/or lower Analog Aux device systems,

Conductivity, DO, ORP, pH and/or Temperature systems. High temperature, high

and low pH and high and low conductivity can also be configured for a timed set

point mode - see “Timed and Normal control mode” information included in spe-

cific sensor setup sections that follow. Configure high and/or low alarm systems

to activate peripheral devices and/or to provide alarm notification messages when

alarm condition(s) exist. Configure values outside of acceptable operating control

range as high and low alarm values.

Set Point Control Values

Range low and Range high values set the acceptable operating range for the water

system. Control system(s) do not become active when value(s) are within acceptable

control range. See figure 4.7 - page 135. Enabled control relays energize when the

sensor reports a value ≤ the range low control value or ≥ range high control value.

Control relays are typically wired to devices used to raise or lower certain water

parameters in order to maintain the system as close to set point as possible. Control

relays remain energized until the sensor reports a value ≤ set point (when dosing

down to lower specific water parameter) and/or ≥ set point (when dosing up to

raise specific water parameter). Exception - high temperature, high and low pH and

high and low conductivity may also be configured for a timed set point mode - see

“Timed and Normal control mode” information included in specific sensor setup

sections that follow. Sensor low control range starts at -1 significant digit below set

point value. High control range starts at +1 significant digit above set point value.

Figure 4.7 shows how the 5200A operates based on a set point control sensor setup

configuration. In the example provided DO range low and range low 2 devices will

energize at ≤ 5.0 and ≤4.0 respectively. The alarm low system becomes active at ≤3.5.

Configuring the 5200A

Range #1

Low

Alarm

ON

High

Alarm

ON

High

Range #1

ON

Optimal

S

etpoint

Value

Lower

Parameter

Values

High

Alarm

Setpoint

Low

Alarm

Setpoint

Higher

Parameter

Values

DO Range #2

High

Range #2

ON

Low

Range #1

ON

Low

Range #2

ON

Acceptable
Operating
Range

Temperature have 1 Control Range

DO has 2 Control Ranges

3.5 mg/l 4.0 5.0 8.1 8.5

8.7 8.9

Figure 4.7

Notes:

- Changing control method between PID/PWM and Set Point may result

in invalid set point, control and alarm value configurations. Always

reconfirm control menus configuration when control mode is changed.

- DO systems have two control ranges. They are referred to as Range Low

& Range Low 2 and Range High & Range High 2.

- Control systems remain active when the user is in 5200A menus.

- Control icons display at the Run Screen when relays are energized - see

Icons - page 70.

- Press

at Run Screen to display active sensor and aux system relays.

- Press

at Run Screen to display events created by control and alarm

relays - see Event Log - page 107.

- Control relays are not energized if the 5X00 DO sensor value reads “Over”

or “Under.”

- “Over” and “Under” are not displayed for aux temperature systems when

temperature values are outside the temperature operating range - see

temperature range specification - page 14 Temperature system controls

will not operate correctly when values are outside the temperature op-

erating range.

- See Wire Relays - page 46 for information on wiring control output

devices.

Set Point Alarm Values

Alarm values set the acceptable control range. Alarm system(s) do not become

active when values are within acceptable control range. See figure 4.7 - page 135.

Enable and configure system alarm(s) to activate alarm output devices(s), view