A Reset Clean Probe message will appear at the Msg’s Menu if the clean probe timer

expires. Selecting “Yes” at “Are you sure?” resets the clean probe timer to the user

configured number of days.

Note:

- User configured sensor and relay names do not appear in log menus.

Calibration Logs

Calibration logs store sensor calibration data. Calibration logs are date and time

stamped and contain sensor and calibration information for completed calibra-

tions.

Certain calibration data is not available at the 5200A. Use AquaManager to view

and save calibration data to the PC. In addition to sensor system, time, date, and

calibration method, AquaManager calibration logs include user input 1, user input

2, and the original and new slope/offset information when applicable.

Event Logs

Control, Alarm, and Timer relays assigned to 5200A systems create events when

they energize and de-energize. In addition, AC power fail and ethernet failure to

initialize errors create events. These events are recorded in the Event log. Events

identify the 5200A system, are time stamped, and record relay on and off times.

The General Alarm does not create an event or control message unless a General

Alarm relay is enabled.

Notes:

- Enable event logging in the System menu. See Event Logging - page 104

for additional information.

Sensor Logs

Enabled systems record sensor data based on data logging menu configuration.

Sensor logs identify sensor system, are date and time stamped, and record sensor

value at user configured intervals.

Notes:

- Data logging must be enabled in the System Menu. See Data Logging -

page 102 for additional information.

Svr Hold Menu

When servicing sensors or doing tank maintenance, it may be necessary to shut

off all systems (including relays) because sensor values will not be accurate. Press

at Run Screen to enter Service Hold menu. When in Service hold, all relays

are de-energized until the Service Hold time expires. The Service Hold time is user

selectable. Enable and configure service hold time at Menus → System→Service

Hold - page 124. After service hold time expires or is cancelled, sensor and aux

systems are reset causing relays to energize/de-energize based on the system value/

Configuring the 5200A

Configuring the 5200A