Service hold, Version, E 129 – YSI 5200A User Manual

Page 65

Advertising
background image

YSI 5200A

129

YSI 5200A

128

1.

Press

to toggle relay on.

2.

Press

to toggle relay off.

3. Press

to save configuration and return to Run Screen.

Note:

- Timer and control icon(s) remain on the Run Screen until the system is

reset. Example - the timer icon remains at the systemwide location until

the end of the timer cycle that was stopped when the relay test menu was

entered even though that relay is not energized. To reset systems after

exiting the Relay Test menu, power cycle the 5200A.

Service Hold

Menus → System → Service Hold

Pressing

at the Run Screen puts the 5200A into Service Hold. The Service

Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any

active relays and disables any sensor, aux, and timer system relays from energizing

regardless of value or condition. 5200A alarm emails are not sent during service

hold. After the service hold time expires or is cancelled, sensor systems are reset

after 30 seconds. Timer systems are not reset. See Service Hold Menu - page 169

for additional information.

Configuring the 5200A

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. Enable and configure submenus.
4. Press

to save configuration and return to Run Screen.

Note:

- Place the 5200A in service hold as instructed in step 3 of calibration checks

in Maintenance, page183

Version

Menus → System →Version

The Version menu contains information about the 5200A including: display and

control board firmware code version, firmware compile date and time, control board

temperature, AC and DC voltage and model number.

Notes:

- Firmware version information is required when contacting YSI sup-

port.

- The clock chip temperature range starts at 0oC. Values < 0oC are not valid

numbers.

- Firmware date and time compilation are displayed below the version

number.

- The normal operating temperature range will be a few degrees above

ambient.

- AC powered units will run at a higher temperature than DC powered

units.

- To view current AC and DC power values press

to return to the

Run Screen and then return to the menu. Power information does not

refresh when the Version menu is open.